Mercury calibrator (mercury vapor dosing device) GA-7300 is a small and handy device operating with the conventional permeation source principle.

Ga-7300 enables both single-point and multipoint calibration, selectable from one picogram to 1000 pg with sampling time 5 min. (Upper range of the injection is proportional to the sampling time used by the analyzer). „Zero picogram dose“ effectively supplies zero-air into the analyzer.

Programmable operation modes: bypass, dosing against zero air and dosing with standard addition.

Calibrator conveniently connects to the analyzer with a standard internet cable. It can be connected to the analyzer anytime. While disconnected, it continues operating on its own, in stand-by.

 

Specifications

Precision of the “zero” (blank value) < 0.2 pg Hg
Programmable dose (at the analyzer sampling time 1 min) 1 – 200 pg Hg
Programmable dose (at the analyzer sampling time 10 min) 1 – 2000 pg Hg
Mercury vapor permeation source temperature control precision 0.06 oC
Air flow rate used for internal system flushing (adjustable) 0.1 – 1 L/min
Communication with the analyzer Internet cable
Time for the device to get conditioned after the prolonged storage 6 hours
Feeding voltage 220 VAC, 110 VAC 12 VDC
Operation ambient temperature range 0 – +40oC
Consumed power, average < 15 W
Consumed power, maximum (at the start up) 40 W
Dimensions (W*D*H) 26 * 35 * 8.5 cm
Weight 4 kg